Drager Babylog 8000 Plus Ventilator

Drager Babylog 8000 Plus Ventilator

The Dräger Babylog 8000 Plus Ventilator is a cornerstone of modern neonatal care, providing the specialized and meticulous support needed to care for neonates with respiratory challenges, ensuring their safe and effective treatment during critical early stages of life.

Questions About This Product?

Our representatives are waiting to answer any questions you may have.

Features:

The Dräger Babylog 8000 Plus Ventilator is a specialized neonatal ventilator designed to provide gentle and precise ventilation to the smallest and most vulnerable patients—newborns and premature infants. This highly advanced ventilator is specifically tailored for neonatal intensive care units (NICUs), where the need for accurate and sensitive respiratory support is crucial.

Key Features of the Dräger Babylog 8000 Plus Ventilator

  • Sophisticated Ventilation Modes: The Babylog 8000 Plus offers a range of ventilation modes, including Pressure Control, Volume Guarantee, and High Frequency Oscillatory Ventilation (HFOV). These modes ensure that even the smallest patients receive the gentlest and most effective respiratory support.
  • Integrated Pulmonary Monitoring: Equipped with advanced monitoring capabilities that allow healthcare providers to track critical respiratory parameters such as tidal volume, peak inspiratory pressure, and oxygen concentration, helping to tailor the ventilation to the patient’s specific needs.
  • Designed for Neonates: Features such as synchronized ventilation and non-invasive support options help reduce the invasiveness of mechanical ventilation and promote spontaneous breathing in neonates.
  • User-Friendly Interface: Comes with a clear, intuitive interface that simplifies adjustments and monitoring, enabling quick responses to the delicate changes in a neonate’s condition.
  • Safety and Alarm Systems: Includes a comprehensive set of alarms designed to alert staff to potential issues quickly, ensuring high levels of safety for these vulnerable patients.
  • Compact and Mobile: Its compact design makes it suitable for use in the space-constrained environment of a NICU and facilitates easy transport within the hospital if needed.

Applications in Medical Care

  • Neonatal Intensive Care Units (NICUs): Primarily used in NICUs to provide vital respiratory support for premature infants and newborns who require assistance due to respiratory distress syndrome (RDS), congenital abnormalities, or other conditions impacting their ability to breathe.
  • Specialized Pediatric Care: Also applicable in settings that require precise and gentle ventilation for small children outside of the typical newborn phase.

Specifications

Controls

  • Conventional ventilation: continuous flow, pressure-limited, time-cycled
  • Triggered ventilation: SIMV, SIPPV, PSV1), leak adapted
    • Trigger: Flow/volume trigger, leak adapted
    • Trigger delay: approx. 40-60 ms
  • High Frequency Ventilation: CPAP+HFV, IMV+HFV; Frequency: 5 to 20 Hz
  • Volume Guarantee Ventilation: SIMV+VG, SIPPV+VG, PSV+VG
  • Oxygen mixer loss (bleed flow): 0 (zero) L/min

Settings

  • Inspiratory oxygen concentration: 21-100 vol.% oxygen
  • Peak inspiratory pressure: 10 to 80 mbar
  • PEEP/CPAP: 0 to 25 mbar
  • Maximal frequency: 200 bpm
  • Inspiratory time: 0.1 to 2 seconds
  • Expiratory time: 0.2 to 30 seconds
  • Inspiratory flow: 1 to 30 L/min
  • Base Flow (VIVE): 1 to 30 L/min

Monitoring

  • Flow monitoring: at the Y-piece, integrated
  • Volume monitoring: at the Y-piece, integrated
  • Lung function monitoring: compliance, resistance C20/C
  • Method: linear regression analysis
  • FiO2 monitoring: integrated
  • Real-time curves: Flow and pressure, integrated
  • Inspiratory oxygen concentration: 21-100 vol.% oxygen
  • Peak pressure: to 99 mbar
  • Mean airway pressure: to 99 mbar
  • Graphic trends: 6 parameters, integrated
  • Logbook: record of up to 100 alarms

Specifications:

Specification Value
Conventional ventilation continuous flow, pressure-limited, time-cycled
Triggered ventilation SIMV, SIPPV, PSV1), leak adapted
High Frequency Ventilation CPAP+HFV, IMV+HFV; Frequency: 5 to 20 Hz
Volume Guarantee Ventilation SIMV+VG, SIPPV+VG, PSV+VG
Oxygen mixer loss (bleed flow) 0 (zero) L/min
Scroll to Top